Comprehensive Visible Imaging Quality Test System ZYP-1630L-VIS
The ZYP visible imaging quality test system combines a collimator, visible source, targets, and other hardware to objectively assess complete visible systems. Targets, name | quantity | specification | application: crosshair | 1 | Φ30-0.3 mm | field and distortion; USAF 1951 resolution | 1 | 0-7 | system resolution; star | 5 | source-written 0.1 mm/0 .2 mm/0 .01mm0.02mm/0 .05mm | system image quality; Porro | 1 | Φ30 mm | focal length; contrast | 5 | 100%, 50%, 33.77%, 15%, 5% | minimum resolvable contrast (MRC); knife edge | 1 | Φ30-R20 mm | modulation transfer function (MTF). Custom options include collimator aperture and focal length, integrated laser or LED sources, target specifications, and rotary stage ranges.
Overview
Specifications
| Parameter | Value |
|---|---|
| Collimator - System Configuration | Off-axis Newtonian reflector |
| Collimator - Focal Length | F 1600mm |
| Collimator - Primary Mirror Material | Quartz |
| Collimator - Effective Aperture D | D 300mm |
| Collimator - Field of View | 1 ° |
| Collimator - Coating | Aluminum+Silicon Protection |
| Collimator - Single-Mirror Reflectance [@0.4 um-14um] | ≥85%@0.4 um-14um |
| Collimator - Operating Wavelength Band | 0.4 um-14um |
| Collimator - On-Axis Field Wavefront Aberration [@632.8 nm] | RMS≤λ/15@632.8 nm |
| Collimator - Parallelism | ≤5″ |
| Halogen Integrating Sphere Light Source - Output Port Diameter | Φ50mm |
| Halogen Integrating Sphere Light Source - Integrating Sphere Inner Diameter | Φ200mm |
| Halogen Integrating Sphere Light Source - Output Uniformity | ≥98% |
| Halogen Integrating Sphere Light Source - Spectral Range | 380nm~2500nm |
| Halogen Integrating Sphere Light Source - Color Temperature | 2856K |
| Halogen Integrating Sphere Light Source - Illuminance Range | 30-3000lux |
| Target Wheel Assembly - Number of Stations | 8 stations |
| Target Wheel Assembly - Control Method | Digital Motorized |
| Target Wheel Assembly - Replacement Method | Manual Replacement |
| Target Wheel Assembly - Positioning Repeatability | 36″ |
| Target Wheel Assembly - Rotary Platform - Drive Type | Motorized |
| Target Wheel Assembly - Rotary Platform - Adjustment Dimensions | Two-Dimensional (Pitch, Horizontal) |
| Target Wheel Assembly - Rotary Platform - Adjustment Range | Pitch ±15°; Horizontal 360° |
| Target Wheel Assembly - Rotary Platform - Resolution | Pitch 0.005°; Horizontal 0.005° |
| Target Wheel Assembly - Rotary Platform - Load Platform Dimensions | 200mm ×200mm |
| Target Wheel Assembly - Rotary Platform - Load Weight | 40kg |
| Control System (Control Cabinet) - Industrial Computer Configuration | Memory 4GB, Disk Space 1TB |
| Control System (Control Cabinet) - Data Interface | Standard Ethernet and USB Interfaces |
| Control System (Control Cabinet) - Hardware Control Software | Control of Visible Light Source, Rotary Table Assembly and Visible-Light Camera |
| Control System (Control Cabinet) - Image Analysis Software | Analysis and Calculation of MTF, MRC, Focal Length, Distortion, Optical Axis, Field of View, etc. |
| Control System (Control Cabinet) - Operating Environment | Windows 10/11 64-Bit, CPU≥Intel i5, Memory ≥8GB, Disk ≥500GB |
| Control System (Control Cabinet) - User Interface | Simplified Chinese Visualization, Supports Parameter Settings, Process Editing and Real-Time Data Display |
| Control System (Control Cabinet) - Data Processing | Automatic Acquisition, Analysis and Storage (Excel/PDF), Supports Historical Data Queries |
| Control System (Control Cabinet) - Report Generation | Custom Templates Automatically Generate Test Reports with Parameters, Results and Charts |
Datasheet
Please provide your email to download the datasheet.
Invalid email format. Please try again.
Supplier Information
Loading...
Name
Phone
Organization
Use Case

Database Platform