VOC Formaldehyde Emission Test Climate Chamber XB-OTS-U1000
The formaldehyde climate chamber measures formaldehyde emissions and was developed according to Chinese standards GB18580-2017 and LY/T1612-2004, US standard ASTMD6007-2, European standard EN717-1, international standard ISO16000-9 and German standard DIN EN 16516-2018. Formaldehyde emissions are an important indicator of wood-based panel quality, related to environmental pollution and human health effects. The 1m3 formaldehyde-emission climate-chamber method is a widely used standard test for indoor decoration and finishing materials. It simulates the indoor climate so that results closely reflect real conditions and are authentic and reliable.
Overview
Specifications
| Parameter | Value |
|---|---|
| Inner Chamber Volume | (1±0.02)m³ |
| Chamber Temperature Range [Temperature Deviation ±0.5°C] | (15—40)°C( Temperature Deviation ±0.5°C) |
| Chamber Humidity Range [Adjustment Precision: ±1.5%R.H.] | (30—80)%R.H.(Adjustment Precision: ±1.5%R.H.) |
| Air Exchange Rate [Accuracy 0.05 times /h] | (0.2-2.0) times / hours ( Accuracy 0.05 times /h) |
| Pressure difference [Slight positive pressure] | 10±5Pa Slight positive pressure |
| Airflow Velocity [Continuously Adjustable] | (0.1—0.3)m/s(Continuously Adjustable) |
| Sampler Pumping Speed [Adjustment Precision: ±5%] | (0.25—2.5)L/min(Adjustment Precision: ±5%) |
| Sealing-Gas Leakage [With an Overpressure of 1000Pa] | Less Than 10-3×1m3/min |
| Sealing-Inlet/Outlet Gas Flow Difference [With an Overpressure of 1000Pa] | Less than 1% |
| Overall dimensions [W*D*H] | Intelligent version (W1100×D1760×H1820)mm |
| Power supply | 120V 16A 60HZ |
| Power - Rated Power | 3KW |
| Power-Operating Power | 2KW |
| Background Concentration Control [Formaldehyde] | Formaldehyde Concentration ≤0.006mg/m³ |
| Thermal insulation | Climate-Chamber Walls and Door Must Provide Effective Thermal Insulation |
| Noise | Climate-Chamber Operating Noise No Greater Than 60dB |
| Continuous Operating time | Climate-Chamber Continuous Operating Time At Least 40 Days |
| Humidity-Control Method | Intelligent combined wet/dry-bulb and dew-point humidity control uses wet/dry-bulb control to reach the set humidity within one hour, then automatically adds dew-point control for stable humidity. Fluctuation is generally no more than 1.5%, well below the required 3%, without water droplets on chamber walls. |
| Control System | A touchscreen forms the operator interface with a German Siemens PLC control system. It directly sets and digitally displays chamber temperature, relative humidity, temperature/dew-point compensation, dew-point/temperature deviation and operating time, and automatically records and plots control curves. |
| Cooling system | Original Imported Italian Aspera Refrigeration Compressor |
| Safety protection System | Refrigeration-system high-pressure and overcurrent protection; heater overtemperature protection; motor overcurrent protection; water shortage protection; and air-pump overheating, overcurrent and overpressure protection. |
| Applicable Standards | GB18580-2017 formaldehyde emission limits for wood-based panels and products used in interior decoration; GB/T39600-2021 formaldehyde emission grading for wood-based panels and products; GB18587-2001 hazardous substance emission limits for carpets, carpet underlays and adhesives; GB17657-2022 physical/chemical testing of wood-based and decorative panels; EN 717-1 chamber determination of formaldehyde release from wood-based panels; ASTM D6007-02 small-chamber measurement of formaldehyde concentrations in emissions from wood products; LY/T1612-2004 1m³ climate chambers for formaldehyde emission testing; GB50325-2020 indoor environmental pollution control for civil buildings; DIN EN 16516-2018 assessment of hazardous substance release from construction products. |