HR2000+ High-Resolution Spectrometer
HR2000+ High-Resolution Spectrometer; The HR2000+ High-Resolution Spectrometer uses a high-resolution optical bench, a 2MHz analog-to-digital (A/D) converter, programmable electronics, a 2048-pixel linear CCD detector, and a high-speed USB2.0 interface. This revolutionary spectrometer not only features a high sampling speed, but can also achieve an optical resolution of 0.035nm (FWHM). When the HR2000+ spectrometer is connected to a computer via the USB2.0 interface, it can acquire and store spectra at a rate of 1000 full-spectrum scans per second, with an optical resolution of 0.035 nm (FWHM). The HR2000+ is very suitable for monitoring chemistry, biochemistry, and other applications requiring fast response as well as fields requiring good spectral characteristics; the HR2000+ integrates a high-resolution optical bench, a powerful 2-MHz A/D converter, programmable electronics, a 2048-pixel CCD array detector, and a USB2.0 port. When the spectrometer is connected to a PC via the USB port, this innovative combination allows the user to capture and store one 0.35 nm (FWHM) full-spectrum scan into memory every millisecond.; The HR2000+ onboard programmable microcontroller provides great flexibility in controlling the spectrometer and accessories. The module has 10 user-programmable data I/O lines that can be connected to other devices, and a pulse generator to trigger other devices. During integration, the user can use the I/O points to light lamps, stop/start a program, and send messages/alarms. The operating parameters of the spectrometer can be controlled through software. The wavelength calibration coefficients that are unique to each spectrometer are programmed into the spectrometer's onboard memory.; By using the Ocean Optics optical bench with high resolution, a 25.4-mm diameter, a 101.6-mm focal length, and a symmetric cross Czerny-Turner optical design, the HR2000+ achieves an optical resolution of 0.035 nm. The HR2000+ response range is 200-1100 nm, but the specific range and resolution depend on your grating and entrance choices. Users can choose from 14 gratings and 6 slit sizes. Ocean Optics offers hundreds of accessories, allowing users to configure special application systems.; Plug-and-play operation: the HR2000+ connects to a PC, PLC, or other embedded system via USB2.0 or an RS-232 serial port. In serial mode, the HR4000 requires an additional 5-volt power supply (not included). The parameters unique to each spectrometer are programmed and stored in the system's memory chip, and can be read very conveniently by the spectrometer operating software.; Users usually require the spectrometer to suit their special needs, so the HR2000+ spectrometer can be configured with an optical bench according to your application requirements. You can choose the slit size, detector, filter, grating, and so on. Our application engineers can help you.; The HR4-BREAKOUT is a passive module that provides interfaces for the various functions of the HR2000+. The BREAKOUT box allows many kinds of connections to the spectrometer, such as: external trigger, GPIO, light source, RS-232, and analog input/output.
Overview
Specifications
| Parameter | Value |
|---|---|
| Dimensions | 148.6 mm × 104.8 mm × 45.1 mm |
| Weight | 570 grams |
| Detector | Sony ILX511 linear silicon CCD-array |
| Detector range | 200-1100 nm |
| Pixel | 2048 pixels |
| Pixel size | 14 μm × 200 μm |
| Pixel well depth | ~62,500 electrons |
| Sensitivity [400 nm] | 75 photons/count |
| Sensitivity [600 nm] | 41 photons/count |
| Design | f/4, Symmetrical crossed Czerny-Turner |
| Focal Length | 101.6 mm input and output |
| Entrance | 5 μm, 10 μm, 25 μm, 50 μm, 100 μm or 200 μm wide slits or fiber (no slit) |
| Grating Selection | 14 different gratings, UV through Shortwave NIR |
| HC-1 grating option | provides 200-1100 nm range |
| Detector Collection Lens Selection | Yes, L2 |
| OFLV Filter Selection | OFLV 200-1100 nm |
| Other Platform Filter Selection | Longpass OF-1 filters |
| Calibration and Focusing Mirror | Standard or SAG+UPG-HR |
| UV Upgrade Window | Yes, UV2 |
| Fiber optic connector | SMA 905 to 0.22 numerical aperture single-strand optical fiber |
| Wavelength Range | Grating dependent |
| Optical resolution | ~0.035-6.8 nm FWHM |
| Signal-to-noise ratio | 250:1 (at full signal) |
| A/D resolution | 14 bit |
| Dark Noise | 3.2 RMS counts |
| Dynamic range | 2 x 10^8 (system); 1300:1 for a single acquisition |
| Integration time | 1 ms to 65 s (20 s typical max.) |
| Polarization [600 nm] | <0.05 % |
| Polarization [435 nm] | <0.10 % |
| Linearity correction | >99.8 % |
| Power consumption [5 VDC] | 450 mA @ 5 VDC |
| Data transfer rate | Full scans to memory every 1 ms with USB 2.0 port, 15 ms with USB 1.1 port |
| Input/Output | Yes, 10 onboard digital user-programmable GPIOs |
| Analog channel | One 13-bit analog input; one 9-bit analog output |
| Auto clear | No |
| Compatibility | Yes, HR4-BREAKOUT |
| Trigger modes | 4 modes |
| Filtering function | Yes |
| Delay characteristics | No |
| Connector | 30-pin connector |
| Operating System | Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
| Computer interface [480 Mbps;RS-232 (2-wire) @ 115.2 K baud] | USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud |
| Peripheral interface | SPI (3-wire); I2C inter-integrated circuit |