LabM-I-4
2.4–4.3µm Detective range ,5mhz Bandwidth Above HgCdTe Programming laboratories Infrared Detection module , Gain Adjustable , With optical impregnated photo-detector LabM-I-4 is A basis for... HgCdTe The heterogeneity is optically impregnated. Photovoltaic The detector's lab. Infrared Detection module , Integration and resistance 、 Programmable Preamplifier 。3° Wedge Sapphire Window prevents unnecessary interference 。 Need a match program “ Smart ”VIGO Thermal electricity cools. Controller PTCC-01( Separately sold ) and smart Manager Software ( Free software ) Use 。LabM-I-4 Modules are applicable to all categories MWIR Pilot development phase in application , The module provides flexible system design for different user needs 。 Specification parameters Spectro-response Features and applications
Overview
Specifications
| Parameter | Value |
|---|---|
| Minimum Cutoff Wavelength λcut-on (10%) [Parameters (Ta = 20°C, Default Settings )] | 2.4±0.5 µm |
| Peak Wavelength λpeak [Parameters (Ta = 20°C, Default Settings )] | 3.5±0.5 µm |
| Optimization Wavelength λopt [Parameters (Ta = 20°C, Default Settings )] | 4 µm |
| Highest Cutoff Wavelength λcut-off (10%) [Parameters (Ta = 20°C, Default Settings )] | 4.3±0.3 µm |
| Detection rate D*(λpeak, 25 kV/A), cm·Hz^(1/2) [Parameters (Ta = 20°C, Default Settings )] | ≥2.0×10¹⁰ |
| Detection rate D*(λopt, 25 kV/A), cm·Hz^(1/2) [Parameters (Ta = 20°C, Default Settings )] | ≥1.2×10¹⁰ |
| Output Noise Density vₙ (10 MHz), nV/Hz^(1/2 ) [Parameters (Ta = 20°C, Default Settings )] | ≤300 |
| Voltage Responsivity Rv (λpeak, 25 kV/A) [Parameters (Ta = 20°C, Default Settings )] | ≥3.5×10⁴ |
| Voltage Responsivity Rv (λopt, 25 kV/A) [Parameters (Ta = 20°C, Default Settings )] | ≥2.3×10⁴ |
| Low Cutoff Frequency flo [Parameters (Ta = 20°C, Default Settings )] | DC/10 (adjustable) |
| High Cutoff Frequency fhi [Parameters (Ta = 20°C, Default Settings )] | ≥5M (adjustable) |
| Output Impedance Rout [Parameters (Ta = 20°C, Default Settings )] | 50 Ω |
| Output voltage Vout [Parameters (Ta = 20°C, Default Settings )] | 1 (RL = 50 Ω*)) |
| Output voltage Offset Voff [Parameters (Ta = 20°C, Default Settings )] | max ±20 |
| Photo-sensitivity component Material [Parameters (Ta = 20°C, Default Settings )] | HgCdTe |
| Light-sensitized area A_(O), mm×mm [Parameters (Ta = 20°C, Default Settings )] | 1×1 |
| Window [Parameters (Ta = 20°C, Default Settings )] | wAl2O3 |
| Receive angle Φ [Parameters (Ta = 20°C, Default Settings )] | ~36° |
| Operating ring temperature requirements Ta [Parameters (Ta = 20°C, Default Settings )] | 10 to 30 °C |
| Signal Output Interface [Parameters (Ta = 20°C, Default Settings )] | SMA |
| Electricity and TEC Control Interface [Parameters (Ta = 20°C, Default Settings )] | LEMO (female) ECG.0B.309.CLN |
| Installation hole [Parameters (Ta = 20°C, Default Settings )] | M4 |
| Fan [Parameters (Ta = 20°C, Default Settings )] | Yes. |
Applications
Key Features
Datasheet
Please provide your email to download the datasheet.
Invalid email format. Please try again.
Supplier Information
Loading...
Name
Phone
Organization
Use Case

Database Platform