High-Resolution Spectrometers PG4000

High-Resolution Spectrometers PG4000

Supplier
Ideaoptics
Part Number
PG4000
Category
Spectrometers
Origin
China

The PG4000 high-resolution spectrometer uses a high-resolution optical platform for applications requiring fine spectral resolution, providing high-quality spectral measurement for laser characterization, gas absorption and plasma analysis. The PG4000 can use a specially UV-sensitized deep-UV CCD to extend detection into the deep ultraviolet. PG4000 flag-bit and low-level access technologies improve measurement accuracy while maintaining high-speed measurement, giving the product greater bargaining power. Low-cost fibers are provided for building customized spectral measurement equipment.

Request a Quote Download Datasheet
Save to Favorites Save to Favorites

Overview

The PG4000 high-resolution spectrometer uses a high-resolution optical platform for applications requiring fine spectral resolution, providing high-quality spectral measurement for laser characterization, gas absorption and plasma analysis. The PG4000 can use a specially UV-sensitized deep-UV CCD to extend detection into the deep ultraviolet. PG4000 flag-bit and low-level access technologies improve measurement accuracy while maintaining high-speed measurement, giving the product greater bargaining power. Low-cost fibers are provided for building customized spectral measurement equipment.

Specifications

ParameterValue
Spectral Model - Spacing [Single channel]0.19 nm
Spectral Model - Spacing [Single channel]0.13 nm
Spectral Model - Resolution [Single channel; 10μm slit; typical value with 20% allowable deviation; FWHM]0.75 nm
Spectral Model - Resolution [Single channel; 10μm slit; typical value with 20% allowable deviation; FWHM]0.49 nm
Spectral Model - Resolution [Single channel; 25μm slit; typical value with 20% allowable deviation; FWHM]0.86 nm
Spectral Model - Resolution [Single channel; 25μm slit; typical value with 20% allowable deviation; FWHM]0.57 nm
Spectral Model - Resolution [Single channel; 50μm slit; typical value with 20% allowable deviation; FWHM]1.29 nm
Spectral Model - Resolution [Single channel; 50μm slit; typical value with 20% allowable deviation; FWHM]0.81 nm
Spectral Model - Resolution [Single channel; 100μm slit; typical value with 20% allowable deviation; FWHM]2.44 nm
Spectral Model - Resolution [Single channel; 100μm slit; typical value with 20% allowable deviation; FWHM]1.58 nm
Spectral Model - Resolution [Single channel; 200μm slit; typical value with 20% allowable deviation; FWHM]4.88 nm
Spectral Model - Resolution [Single channel; 200μm slit; typical value with 20% allowable deviation; FWHM]3.16 nm
Spectral Model - Spacing [5 Channel]0.06 nm
Spectral Model - Spacing [5 Channel]0.05 nm
Spectral Model - Spacing [5 Channel]0.04 nm
Spectral Model - Resolution [5 Channel ;10μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.20 nm
Spectral Model - Resolution [5 Channel ;10μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.15 nm
Spectral Model - Resolution [5 Channel ;10μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.14 nm
Spectral Model - Resolution [5 Channel ;25μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.24 nm
Spectral Model - Resolution [5 Channel ;25μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.21 nm
Spectral Model - Resolution [5 Channel ;25μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.22 nm
Spectral Model - Resolution [5 Channel ;50μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.39 nm
Spectral Model - Resolution [5 Channel ;50μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.38 nm
Spectral Model - Resolution [5 Channel ;50μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.35 nm
Spectral Model - Resolution [5 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.81 nm
Spectral Model - Resolution [5 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.77 nm
Spectral Model - Resolution [5 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.75 nm
Spectral Model - Resolution [5 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.68 nm
Spectral Model - Resolution [5 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.58 nm
Spectral Model - Resolution [5 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.57 nm
Spectral Model - Resolution [5 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.50 nm
Spectral Model - Resolution [5 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.36 nm
Spectral Model - Spacing [3 Channel]0.05 nm
Spectral Model - Spacing [3 Channel]0.12 nm
Spectral Model - Resolution [3 Channel ;10μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.19 nm
Spectral Model - Resolution [3 Channel ;10μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.43 nm
Spectral Model - Resolution [3 Channel ;25μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.25 nm
Spectral Model - Resolution [3 Channel ;25μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.53 nm
Spectral Model - Resolution [3 Channel ;50μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.38 nm
Spectral Model - Resolution [3 Channel ;50μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.79 nm
Spectral Model - Resolution [3 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]0.79 nm
Spectral Model - Resolution [3 Channel ;100μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.58 nm
Spectral Model - Resolution [3 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]1.58 nm
Spectral Model - Resolution [3 Channel ;200μm Slit ; typical , Actually Permitted 20% Deviation ;FWHM]3.14 nm
Spectral Performance - Detection Range [Depends on the grating]200~1100 nm,具体视光栅而定
Spectral Performance - Wavelength ResolutionUp to 0.04nm with an 1800-line grating; see the Model Table for details
Spectral Performance - Optical ResolutionUp to 0.12nm (FWHM) with an 1800-line grating and 10μm slit; see the Model Table for details
Spectral Performance - DetectorToshiba, 3648 pixels CCD
Spectral Performance - Higher-Order Diffraction Elimination3 front and 4 rear filter options to eliminate ghost lines in the spectrum
Spectral Performance - Optical BenchFocal length 100mm, f-number 4.5, crossed C-T optical path
Spectral Performance - Integration Time4ms ~ 60s
Spectral Performance - Dynamic RangeBetter than 2000:1, significantly better than comparable products
Spectral Performance - Signal-to-Noise Ratio [At saturation]Better than 300:1 at saturation
Spectral Performance - Stray Light [Measurement wavelength 600nm]< 0.1%, measurement wavelength 600nm
Spectral Performance - Corrected Linearity>99%
External Interfaces - Fiber InterfaceSMA905
External Interfaces - Data TransmissionUSB-B (square port), supporting USB2.0 communication protocol and 480Mbps transfer speed
External Interfaces - Expansion Function Port30 pin expansion port, supporting Ideaoptics BreakOut-Board and BreakOut-Cable for expanded external functions
External Interfaces - External Trigger4 trigger modes: Normal / Level / Edge / Synchronization
Other - Dimensions157 mm³ × 115 mm³ × 41.5 mm³
Other - Weight0.81 kg
Other - A/D Bit Depth16-bit
Other - Power Consumption200mA @ 5VDC
Other - High-Speed Control TechnologyNew spectrum acquisition starts within 4ms after the host issues an integration-time adjustment command
Other - Flag-Bit TechnologyProvides an acquisition-complete flag for host queries

Applications

Industrial and consumer electronics: measurement of core optical parameters such as laser pulse width for research, consumer electronics, medical aesthetics and industrial laser welding. Absorbance measurement: the absorption spectrum is the ratio of the sample spectrum to the reference spectrum, reducing the original spectral signal-to-noise ratio and requiring a high-signal-to-noise spectrometer; a broad spectral range is also needed for more information. Transmission/reflection measurement: diverse samples and complex test conditions require strong system versatility and adaptability. Shendu Guangpu applications in photovoltaics and optoelectronics.

Key Features

High-resolution optical platform: optical resolution up to 0.1nm, with a 100mm focal length and 0.11 numerical aperture balancing resolution and sensitivity without increasing instrument size. EX dual-blaze grating: a more uniform broadband response balances efficiency and addresses higher-order interference, with maximum coverage of 200~1100nm. High-speed control: a new integration time can be set within 1ms, reducing instrument-control time. Excellent signal-to-noise ratio; high optical resolution of 0.1nm; optional specially UV-sensitized CCD; suitable for fine spectral resolution.

Datasheet

Product Datasheet
PDF
Download

Supplier Information

Id
Ideaoptics
View Supplier →
Shanghai Ideaoptics Co., Ltd. is a national-level specialized and sophisticated 'Little Giant' enterprise, a deep-spectrum technology innovator. Devoted to the micro-nano optoelectronic field, it develops intelligent full-spectrum technology integrating photonics and AI, building an internationally leading deep-spectrum platform. The company provides comprehensive solutions from technology to products, including micro spectrometers, angle-resolved spectrometers, and confocal Raman spectrometers, widely applied in IC chips, optical chips, biochips, AR/VR, and advanced displays. Through joint R&D centers with universities, Ideaoptics continuously launches breakthrough products, serving over 3,000 customers and cooperating with 170+ semiconductor, advanced materials, and biomedical enterprises, aiming to advance photonic integration and empower semiconductor manufacturing.

Request a Quote

Our team responds within 24–48 hours
High-Resolution Spectrometers PG4000
Ideaoptics · Spectrometers
Direct from Manufacturer
Your inquiry will be sent directly to the manufacturer
Photonics Source Platform
We handles the inquiry on your behalf
Your information is protected and will never be shared without consent.
Contact Us

Loading...

GetExperimental Plan

Name

Phone

+86

Organization

Use Case

AI Parametric Selection
Smart Match · Precise Recs
Enter Specifications
AI Smart Matching
Get Best Matches
Skip Manual Comparison
Find Best Solutions Fast
Select Now →