iHR-Series Imaging Spectrometer iHR550

iHR-Series Imaging Spectrometer iHR550

Supplier
Dowellphotonics
Part Number
iHR550
Category
Spectrometers
Origin
China
Specification Snapshot
Single-Pixel Size [Parameter Figure 03; Compatible with Synapse™, Symphony® and Syncerity™ Series CCDs; Separate Explanation of Compatible CCDs]
13.5~26 μm
Pixel Count [Parameter Image 03; Compatible CCD Range; Independent Description of Compatible CCDs]
512~2048
Cooling Method [Parameter Image 03 Compatible CCD Range; Independent Description of Compatible CCDs]
TE Cooling to Liquid Nitrogen Cooling
Detector Type [Parameter Image 03; Liquid-Nitrogen-Cooled SymphonyII Detector; Independent Description of Compatible CCDs]
Silicon, InGaAs and Extended InGaAs
Spectral Range [Parameter Image 03; Liquid-Nitrogen-Cooled SymphonyII Detector; Independent Description of Compatible CCDs]
200nm~2.2μm
Focal Length [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]
550 mm
Request a Quote Download Datasheet
Save to Favorites Save to Favorites

Specifications

ParameterValue
Single-Pixel Size [Parameter Figure 03; Compatible with Synapse™, Symphony® and Syncerity™ Series CCDs; Separate Explanation of Compatible CCDs]13.5~26 μm
Pixel Count [Parameter Image 03; Compatible CCD Range; Independent Description of Compatible CCDs]512~2048
Cooling Method [Parameter Image 03 Compatible CCD Range; Independent Description of Compatible CCDs]TE Cooling to Liquid Nitrogen Cooling
Detector Type [Parameter Image 03; Liquid-Nitrogen-Cooled SymphonyII Detector; Independent Description of Compatible CCDs]Silicon, InGaAs and Extended InGaAs
Spectral Range [Parameter Image 03; Liquid-Nitrogen-Cooled SymphonyII Detector; Independent Description of Compatible CCDs]200nm~2.2μm
Focal Length [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]550 mm
Relative Aperture [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]f/6.4
Spectral Range [Parameter Figure 02 Basic Technical Specifications; All Parameters Based on a 1200g/mm Grating at 435nm, except Items Explicitly Specifying Another Wavelength; Using a 1200g/mm Grating]150~1500 nm
Maximum Spectral Coverage [Parameter Figure 02 Basic Technical Specifications; All Parameters Based on a 1200g/mm Grating at 435nm, except Items Explicitly Specifying Another Wavelength; Using Another Suitable Grating Specification]150nm~40μm
Grating Dimensions [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]76 mm × 76 mm
Number of Gratings on the Turret [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]Up to 3
Focal-Plane (Flat-Field) Dimensions [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]30 mm × 12 mm
Resolution [Parameter Figure 02 Basic Technical Specifications; All Parameters Based on a 1200g/mm Grating at 435nm, except Items Explicitly Specifying Another Wavelength; Slit/PMT]0.025 nm
Wavelength Accuracy [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]±0.20 nm
Repeatability [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]±0.075 nm
Spectral Dispersion [Parameter Figure 02 Basic Technical Specifications; All Parameters Based on a 1200g/mm Grating at 435nm, except Items Explicitly Specifying Another Wavelength; This Item Explicitly Specifies @500nm]1.34 nm/mm
Magnification [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]1.1
Stray Light [Parameter Figure 02*; Using a HORIBA Scientific Holographic Grating, Measured 1nm from the Peak Baseline of the 514nm Laser Line]1×10⁻⁵
Scanning Speed [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]160 nm/s
Step Size [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]0.002 nm
Computer Interface [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]High-Speed USB
Length [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item; Independent Technical Table Dimensions]648 mm
Width [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item; Independent Technical Table Dimensions]460 mm
Height [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item; Independent Technical Table Dimensions]193 mm
Optical-Axis Height [Parameter Figure 02 Basic Technical Specifications; All Parameters Based on a 1200g/mm Grating at 435nm, except Items Explicitly Specifying Another Wavelength; Measured from Instrument Bottom]98 mm
Weight [Parameter Image 02 Basic Technical Specifications; All Parameters Given for a 1200g/mm Grating at 435nm, Except Where a Wavelength Is Explicitly Stated for the Item]28 kg
Groove Density3600 g/mm
Linear Dispersion0.16 nm/mm
Grating-Turret Mechanical Range0~500 nm
Spectral Range per Acquisition4 nm
Spectral Range per Pixel0.002 nm
Typical Spectral Resolution0.01 nm
Spectral Range per Pixel0.004 nm
Typical Spectral Resolution0.01 nm
Groove Density2400 g/mm
Linear Dispersion0.53 nm/mm
Grating-Turret Mechanical Range0~750 nm
Spectral Range per Acquisition14 nm
Spectral Range per Pixel0.007 nm
Typical Spectral Resolution0.02 nm
Spectral Range per Pixel0.014 nm
Typical Spectral Resolution0.04 nm
Groove Density1800 g/mm
Linear Dispersion0.81 nm/mm
Grating-Turret Mechanical Range0~1000 nm
Spectral Range per Acquisition22 nm
Spectral Range per Pixel0.011 nm
Typical Spectral Resolution0.03 nm
Spectral Range per Pixel0.021 nm
Typical Spectral Resolution0.06 nm
Groove Density1200 g/mm
Linear Dispersion1.34 nm/mm
Grating-Turret Mechanical Range0~1500 nm
Spectral Range per Acquisition36 nm
Spectral Range per Pixel0.018 nm
Typical Spectral Resolution0.05 nm
Spectral Range per Pixel0.035 nm
Typical Spectral Resolution0.10 nm
Groove Density900 g/mm
Linear Dispersion1.84 nm/mm
Grating-Turret Mechanical Range0~2000 nm
Spectral Range per Acquisition49 nm
Spectral Range per Pixel0.025 nm
Typical Spectral Resolution0.07 nm
Spectral Range per Pixel0.048 nm
Typical Spectral Resolution0.14 nm
Groove Density600 g/mm
Linear Dispersion2.83 nm/mm
Grating-Turret Mechanical Range0~3000 nm
Spectral Range per Acquisition76 nm
Spectral Range per Pixel0.038 nm
Typical Spectral Resolution0.11 nm
Spectral Range per Pixel0.074 nm
Typical Spectral Resolution0.22 nm
Groove Density300 g/mm
Linear Dispersion5.75 nm/mm
Grating-Turret Mechanical Range0~6000 nm
Spectral Range per Acquisition154 nm
Spectral Range per Pixel0.078 nm
Typical Spectral Resolution0.23 nm
Spectral Range per Pixel0.150 nm
Typical Spectral Resolution0.45 nm
Groove Density150 g/mm
Linear Dispersion11.58 nm/mm
Grating-Turret Mechanical Range0~12000 nm
Spectral Range per Acquisition309 nm
Spectral Range per Pixel0.156 nm
Typical Spectral Resolution0.47 nm
Spectral Range per Pixel0.301 nm
Typical Spectral Resolution0.90 nm

Key Features

iHR Means Imaging High Resolution,Excellent Imaging/Spectral Resolution;Unique Asymmetric Czerny-Turner rather than Common Rectangular Symmetric Path,Optimizes Imaging,Eliminates Secondary Diffraction,Maximizes Throughput;Re-Diffracted Light Returning to Grating Causes Dim Background/Baseline Rise/Lower Sensitivity across Bands,Especially Symmetric Designs;Computer-Simulated Optimized Optic Positions Precisely Eliminate It;With CCD,Toroidal Mirrors Correct Astigmatism,Tangential Resolution/Sagittal Imaging Optimization,Adjustable CCD Prioritizes Imaging or Resolution;Asymmetric Path/On-Axis Grating Scan Reduces Coma/Other Aberrations for Wider Flat Field;Toroidal Focus Mirror Larger than Collimator Avoids Edge Throughput Loss/Vignetting;Standard/Custom Single/Dual Inputs/Outputs,Grating Turret,Single/Multichannel Detectors,Sample Chamber,Fiber/Adapters,Filters,Shutters,CCD Flanges;SynerJY®,Labspec Raman,SDK,LabVIEW VIs;Seamless One-Piece Casting for Strength/Thermal Stability,Optical Chamber Separate from Electronics,Strict Aging Cycles/Repeated Drive Tests;Synapse™/Symphony®/Syncerity™ CCDs,Front/Open-Electrode through Back/Deep-Depletion by Experiment

Datasheet

Please provide your email to download the datasheet.

Your name
Email Address

Invalid email format. Please try again.

Send Request

Supplier Information

Do
Dowellphotonics
View Supplier →
Starting from basic and simple optical spectroscopy and imaging modules to multi-functional and complex spectroscopy and imaging system solutions, Dowell Photonics has been committed to providing customized, personalized services and experiences. With exquisite technology, we design clever and flexible optical modules, making your experimental testing, R&D, and production no longer limited by instrument functions. We help you build, expand, and upgrade multi-functional spectroscopic testing systems to meet your special needs!

Request a Quote

Our team responds within 24–48 hours
iHR-Series Imaging Spectrometer iHR550
Dowellphotonics · Spectrometers
Direct from Manufacturer
Your inquiry will be sent directly to the manufacturer
Photonics Source Platform
We handles the inquiry on your behalf
Your information is protected and will never be shared without consent.
Contact Us

Loading...

GetExperimental Plan

Name

Phone

+86

Organization

Use Case

AI Parametric Selection
Smart Match · Precise Recs
Enter Specifications
AI Smart Matching
Get Best Matches
Skip Manual Comparison
Find Best Solutions Fast
Select Now →