Experimental/Processing System Connecting a Twelve-Glove-Port Glovebox to a Single-Station Glovebox with Heated Transfer Chamber
A double-sided glovebox with twelve glove ports integrates equipment into a single large chamber without physical internal partitions, allowing large equipment such as robot tracks and samples to move freely. Operators on both sides can access most of the interior simultaneously for transfer of complex parts and collaborative assembly. A heated antechamber is described as key to long-term sub-ppm ultrahigh purity: substrates, wafers, tools and packaging are baked at high temperature before entering the main chamber to remove moisture and organics adsorbed on surfaces and in pores. In a single-workstation glovebox, one circulation/purification system maintains uniform moisture and oxygen concentrations for preparation, coating and drying of highly sensitive samples. The system can accommodate a complex process chain that would otherwise require transfers between separate ultraclean rooms or gloveboxes. It is presented as an engineering solution for complex assembly and manufacturing without air exposure, serving advanced research with stringent reliability, purity and integration requirements. Nonstandard twelve-port purging glovebox: double-sided, two-workstation chamber, dimensions stated as 1500m X 1000mm X 1500mm. Inclined working face with removable 8 mm safety-tempered front glass, resistant to wear and corrosion with good light transmission; 3/8-inch-thick OMEGA seal. Two internal shelving levels are optional. Standard double-sided, two-workstation glovebox: dimensions 1200mm X 1000mm X 900mm; inclined working face with removable 8 mm safety-tempered front glass, resistant to wear and corrosion with good light transmission; 3/8-inch-thick OMEGA seal. Two internal shelving levels are optional. Large antechamber: cylindrical 304 stainless steel, connected to the glovebox, with an easy-to-operate shaft-lock door; 360 mm diameter and 600 mm long, piston door and polished sliding tray. Small antechamber: 150 mm diameter and 300 mm long, with manual evacuation/refill valves. Automatic purging supports atmosphere changes and rapid purging after accidental water or oxygen ingress. The connection from the automatic purging valve to the chamber piping manifold uses securely attached KF25 steel-wire-reinforced plastic tubing, with an external KF25 barbed fitting. An automatic pneumatic electroplated panel is specified. Users set the chamber gas-purging duration and touch Start on the touchscreen; purging stops automatically at the set time. Vacuum-pump start/stop timing control is added. Oxygen analyzer: imported U.S. Avcray electrochemical instrument, neither domestically manufactured nor assembled domestically; third-party data are stated to support reliability. Alternative: GE oxy.IQ™ Oxygen Transmitter. Moisture analyzer: imported British MICHELL, selected according to the customer's operating environment, neither domestically manufactured nor assembled domestically; third-party data are stated to support reliability. Alternative: GE VeriDri™ Dew-Point Transmitter. Analyzer configuration applies to the standard glovebox. The nonstandard twelve-port purging glovebox reserves moisture/oxygen-analyzer and circulation/purification interfaces for convenient later expansion. Solvent purification: a solvent column is connected in series to protect the purification column from organic solvents and remove volatilized solvents. The principal medium is high-quality, high-efficiency activated carbon. The adsorber is charged with 7 KG of washed and dried high-efficiency activated carbon and activated alumina, installed beneath the chamber, with KF40 ports at the top and bottom for user replacement. This solvent-purification configuration applies to the standard glovebox; the nonstandard twelve-port purging glovebox reserves moisture/oxygen-analyzer and circulation/purification interfaces for later expansion. Product certifications stated: ISO9001, CE and UL. Warranty: one year, with lifetime repair service. Application precautions: consult the Danger, Warning, Caution and related sections of the instruction manual.
Overview
Specifications
| Parameter | Value |
|---|---|
| Clean-out glove box - box - size | 1500mX1000mmX1500mm |
| Cleaning glove box - box - structure | Total stainless steel structure (Type304) |
| Cleaning glove box - box - thickness | 3 mm |
| Cleaning glove box - box - volume | 2.2 m³ |
| Cleaning glove box -- box -- front window -- glass thickness, non-stamped glove box | 8 mm |
| Cleaning glove boxes - boxes - front windows - seals - non-stamped gloves | 3/8 inches thick OMEGA seal. |
| Cleaning glove box-box-sketter-stalk-sketter | 8 in |
| Cleaning glove boxes - boxes - gloves - materials | Hard aluminium alloy gloves (combustion-resistant, aluminium anode oxide, electrolytic and organic solvent corrosion resistance) |
| Cleaning glove box - box - glove mouth - calibre | 220 mm |
| Cleaning glove boxes - boxes - gloves - materials | Original butyl rubber imports, United States |
| Cleaning glove box - box - glove - thickness | 0.4 mm |
| Nonstandard Twelve-Glove-Port Purging Glovebox—Chamber—Filter—Pore Size | 0.3 μm |
| Cleaning glove boxes -- boxes -- lighting | The box is equipped with a reaction-proof light-saving light, soft light. |
| Cleaning glove box - box - leak rate | ≤0.009 vol%/h |
| Cleaning glove boxes -- boxes -- shelf [Optional] | Two-tier shelving |
| Cleaning glove boxes - boxes - power panels - holes | 6 孔 |
| Cleaning glove boxes - box - power panels - power supply | 220V±10% 50Hz±10% |
| Cleaning glove box -- box -- interface, non-stamped glove box | Multiple standard KF-40 interfaces set |
| Cleaning glove box -- box -- support frame -- structure | Steel structure |
| Cleaning glove box -- box -- stairwell -- high-bearing foot wheel | 950 mm |
| Cleaning glove box - box - stubble - leg wheel | A thousand-way wheel. |
| Standard double-sided double-space glove box - box size | 1200mmX1000mmX900mm |
| Standard double-sided double-space glove box - box - structure | Total stainless steel structure (Type304) |
| Standard double-sided double-space glove box - box - thickness | 3 mm |
| Standard double-sided double-dot glove box - box size | 1.0 m³ |
| Standard double-sided double-space glove box - box - front window - glass thickness | 8 mm |
| Standard double-sided double-space glove box - box - front window - seal | 3/8 inches thick OMEGA seal. |
| Standard double-sided double-space glove box - box - glove mouth | 8 in |
| Standard double-sided double-space glove box - box - glove mouth - material | Hard aluminium alloy gloves (combustion-resistant, aluminium anode oxide, electrolytic and organic solvent corrosion resistance) |
| Standard double-sided double-space glove box - box - glove mouth - calibre | 220 mm |
| Standard double-dot glove box - box - glove - material | Original butyl rubber imports, United States |
| Standard double-sided double-space glove box - box - glove - thickness | 0.4 mm |
| Standard double-dot glove box - box - filter - aperture | 0.3 μm |
| Standard double-sided double-space glove box - box - lighting | The box is equipped with a reaction-proof light-saving light, soft light. |
| Standard double-dot glove box - box - leakage rate | ≤0.009 vol%/h |
| Standard double-sided double-space glove box - box - shelf [Optional] | Two-tier shelving |
| Standard double-dot glove box - box - power panels - holes | 6 孔 |
| Standard double-dot glove box - box - power panels - power supply | 220V±10% 50Hz±10% |
| Standard double-sided double-space glove box - box - interface | Multiple standard KF-40 interfaces set |
| Standard double-sided double-space glove box - box - support frame - structure | Steel structure |
| Standard double-sided double-space glove box - box - support frame - high footer | 950 mm |
| Standard double-sided double-space glove box - box - support frame - foot wheel | A thousand-way wheel. |
| Transition class -- large transition class -- structure | Cylinder |
| Transition class -- large transition class -- materials | 304 stainless steel |
| Transition class -- large transition class -- door | Axle locks, big bridge door. |
| Transition class -- large transition class -- door | Piston Gate |
| Transition class -- large transition class -- diameter | 360 mm |
| Transition class -- large transition class -- long | 600 mm |
| Transition class -- large transition module -- pallet | Slideable Tray |
| Transition class -- large transition module -- vacuum [Large, small right overboard requires a vacuum belt to be displayed] | <1 bar |
| Transition class -- small transition module -- materials | stainless steel type 304 (1.4301) |
| Transition class - small transition module - thickness | 2 mm |
| Transition class -- small transition class -- diameter | 150 mm |
| Transition class -- small transit class -- long | 300 mm |
| Transition class -- small transition module -- vacuum [Large, small right-hand transit module requires a vacuum belt to be displayed] | ≤1 bar |
| Purge system - valve to box pipe integration | Strongly connected KF25 steel wire plastic tube |
| Cleaning system -- external interface | KF25 tower. |
| Analyzers - oxygen analysers - dimensions [Standard glove box set] | 205 mm(L) x 80 mm(W) x 60 mm(H) |
| Analyzers - oxygen analysers - range measurement [Standard glove box set] | 0-1000 Ppm |
| Analyzer-water analyser-size [Standard glove box set] | 205 mm(L) x 80 mm(W) x 60 mm(H) |
| Analyzers - hydrometric analysers - range of measurements [Standard glove box set] | 0-1000 ppm |
| Solvent purification system - media [Standard glove box set] | High-quality, high-efficiency, active charcoal |
| Solvent purification systems - fillings [Standard glove box set] | High-impact activated carbon and aluminum oxide dryed with water |
| Solvent purification systems - filled quantity [Standard glove box set] | 7 KG |
| Solvent purification system -- location of installation [Standard glove box set] | Bottom of the box. |
| Solvent purification system - interface [Standard glove box set; above and below the column] | KF40 interface |
| Cleaning glove box for non-stamped gloves -- preset interface | Hydrogen analyser interface, cycle purification system interface, etc. |
Key Features
Datasheet
Supplier Information
Loading...